Class II Biosafety Cabinet Series
Product Design Compliance
The product design adheres to China CFDA’s YY 0569 standard and the Class II biosafety requirements of the US NSF/ANSI 49 standard.
Ergonomic & Safety Design
- Front Window: Features a 10° ergonomic tilt angle, optimizing operational comfort and human-machine interaction.
- Double-Layer Cabinet: Enclosed in a surrounding negative-pressure structure to prevent leakage and contamination.
Intelligent Human-Machine Interface
- 7-inch HD Touchscreen: A high-definition LED display enabling intelligent device management.
- Automatic Airflow Compensation: Automatically adjusts fan speed when abnormal airflow is detected.
- Anomaly Monitoring: Continuously monitors system status; triggers audible and visual alarms for faults.
- Scheduled On/Off: Allows presetting of start and shutdown times for automated operation.
Advanced Manufacturing Technology
- Self-Cleaning Structure: The exposed working area uses an R12.5 three-sided seamless rounded corner design, enhancing self-cleaning efficiency.
- Material Durability: Inner surfaces of the working area are constructed from 304 stainless steel for corrosion resistance.
- Safety Glass: The front window adopts anti-shattering tempered glass—optically clear, with no degradation during cleaning or disinfection.
Rigorous Factory Testing
- Cabinet Leakage Test: No leakage detected after pressurizing the cabinet to 500Pa.
- Filter Integrity Test:
- For scannable HEPA/ULPA filters: Leakage rate ≤ 0.01%.
- For non-scannable detection points: Leakage rate ≤ 0.005%.